| Model Number |
1219-28-250 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Date event (=date the issue took place): (b)(6) 2023.( revision ) implantation: (b)(6) 2019.Procedure: total hip after a fracture.Description of issue: luxation of liner, scallops of the liner have broken off.When did the even occur (pre-op/intra-op/post-op)? post-op.Was there any patient consequence? yes , revision surgery on date: (b)(6) 2023.What action was taken to resolve the issue? revision surgery, placed a new liner.How long was the procedure prolonged as a result of the difficulties encountered with the device (minutes)?.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence of the device confirmed the reported allegation.It was observed that interlock feature edge of the liner denoted four anti-rotation device (ard) tabs sheared off from the liner.Based on the observations, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the photographic evidence of the device confirmed the reported allegation.It was observed that interlock feature edge of the liner denoted four anti-rotation device (ard) tabs sheared off from the liner.Based on the observations, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|