• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the blade was fractured and separated from the device tip.The broken tip was not returned.Inspection of the fractured blade revealed signs of gouging and outer coating erosion as a result of activating against solid/hard materials.The results of the investigation determined that the reported event was confirmed for fractured blade.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: too much force or torque applied to instrument, or grasping/pulling; incidental and prolonged activation against solid surfaces, such as bone or plastic; attempting to bend, sharpen, or otherwise alter the shape of the blade.The instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the blade of the device broke inside the patient and that the piece was recovered manually.The device was replaced and the procedure was completed successfully.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key16898794
Report Number0001056128-2023-00014
Device Sequence Number20198477
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public07613327346251
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K161693
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2025
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number14310000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/14/2023
Initial Report FDA Received Date05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-