| Model Number |
IPN920375 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Foreign Body In Patient (2687)
|
| Date of Event |
04/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Medwatch received reports: "on [redacted date], the patient underwent coronary brachytherapy via right femoral arterial access.Later that same day while the provider was attempting to gently remove the femoral access catheter with minimal traction, the head of the catheter and side port fractured off with the catheter shaft still remaining in situ and a rush of brisk arterial blood followed.The provider immediately applied and maintained manual pressure proximally to the arteriotomy site against the femoral head, with he mostasis was achieved.Vascular surgery was stat paged to the patient's bedside as provider continued to hold manual pressure throughout the entire time.The patient was then taken emergently to the or for exploration and removal of retained access catheter.The specific introducer used was a "set introducer super arrow-flex 7fr 11cm hemostatic valve percutaneous sheath integral side port attached 3way stopcock radiopaque tip orange cs/10ea".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
Medwatch received reports: "on [redacted date], the patient underwent coronary brachytherapy via right femoral arterial access.Later that same day while the provider was attempting to gently remove the femoral access catheter with minimal traction, the head of the catheter and side port fractured off with the catheter shaft still remaining in situ and a rush of brisk arterial blood followed.The provider immediately applied and maintained manual pressure proximally to the arteriotomy site against the femoral head, with he mostasis was achieved.Vascular surgery was stat paged to the patient's bedside as provider continued to hold manual pressure throughout the entire time.The patient was then taken emergently to the or for exploration and removal of retained access catheter.The specific introducer used was a "set introducer super arrow-flex 7fr 11cm hemostatic valve percutaneous sheath integral side port attached 3way stopcock radiopaque tip orange cs/10ea".
|
| |
|