ARROW INTERNATIONAL LLC ARROW CATH LAB SHEATH INTRO SET; CATHETER PERCUTANEOUS Back to Search Results
Model Number IPN920375
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Foreign Body In Patient (2687)
Date of Event 04/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Medwatch received reports: "on [redacted date], the patient underwent coronary brachytherapy via right femoral arterial access.Later that same day while the provider was attempting to gently remove the femoral access catheter with minimal traction, the head of the catheter and side port fractured off with the catheter shaft still remaining in situ and a rush of brisk arterial blood followed.The provider immediately applied and maintained manual pressure proximally to the arteriotomy site against the femoral head, with he mostasis was achieved.Vascular surgery was stat paged to the patient's bedside as provider continued to hold manual pressure throughout the entire time.The patient was then taken emergently to the or for exploration and removal of retained access catheter.The specific introducer used was a "set introducer super arrow-flex 7fr 11cm hemostatic valve percutaneous sheath integral side port attached 3way stopcock radiopaque tip orange cs/10ea".
 
Additional Manufacturer Narrative
Qn#(b)(4).
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Medwatch received reports: "on [redacted date], the patient underwent coronary brachytherapy via right femoral arterial access.Later that same day while the provider was attempting to gently remove the femoral access catheter with minimal traction, the head of the catheter and side port fractured off with the catheter shaft still remaining in situ and a rush of brisk arterial blood followed.The provider immediately applied and maintained manual pressure proximally to the arteriotomy site against the femoral head, with he mostasis was achieved.Vascular surgery was stat paged to the patient's bedside as provider continued to hold manual pressure throughout the entire time.The patient was then taken emergently to the or for exploration and removal of retained access catheter.The specific introducer used was a "set introducer super arrow-flex 7fr 11cm hemostatic valve percutaneous sheath integral side port attached 3way stopcock radiopaque tip orange cs/10ea".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CATH LAB SHEATH INTRO SET
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16898797
Report Number9680794-2023-00332
Device Sequence Number5459250
Product Code DQY
UDI-Device Identifier10801902194173
UDI-Public10801902194173
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN920375
Device Catalogue NumberCP-07711
Device Lot Number14F22E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Supplement Date Received by Manufacturer06/06/2023
Initial Report FDA Received Date05/09/2023
Supplement Report FDA Received Date06/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexFemale
Patient Weight73 KG
Patient EthnicityHispanic
-
-