| Model Number |
0475000000 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.The patient was under anesthesia, but no incision was made.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device did not boot up.Probable root cause: software error.Front board issue.Blown fuse.Power board issue.Electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge.Insufficient cybersecurity (cf).The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was cancelled.The patient was under anesthesia, but no incision was made.
|
| |
|