• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION LED LIGHT SOURCE; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, PRECISION LED LIGHT SOURCE; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Model Number 0220220000
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: was hot to the touch.Probable root cause: fan malfunction.Main board malfunction.Led module malfunction.Thermal switch malfunction.Use errors.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, PRECISION LED LIGHT SOURCE
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16899268
Report Number0002936485-2023-00412
Device Sequence Number20176418
Product Code FCW
UDI-Device Identifier07613327051322
UDI-Public07613327051322
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0220220000
Device Catalogue Number0220220000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2023
Supplement Date Received by Manufacturer04/10/2023
Initial Report FDA Received Date05/09/2023
Supplement Report FDA Received Date02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-