| Model Number |
H-1129 |
| Medical Device Problem Codes |
Leak/Splash (1354); Power Conditioning Problem (1474)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the unit was "tripping" the electrical box in the operating room when it was powered on at the start of a procedure.It was also discovered the unit had a "puddle" underneath it when the problem was investigated.No adverse effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6 evaluation codes - updated one device was received for investigation with no cuffs, no hi 2 mounting bracket and no ram horn.After further visual inspection, it was found that the ground wire to the back panel was taken off and placed on the pump bracket which is incorrect.Functional testing occurred and was unable to duplicate the unit tripping the electrical box, however, it did verify that the unit was leaving a puddle underneath the device.A crack returned tube caused this leak underneath the device.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date of 2016-06-12 and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.This issue will continue to be monitored and further actions taken accordingly.
|
| |
|
Search Alerts/Recalls
|