• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; RESURFACING FEMUR, LEFT, SIZE 4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; RESURFACING FEMUR, LEFT, SIZE 4 Back to Search Results
Model Number RF-L004E-01M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2023 where the following eleos implants were explanted: two cemented stem extensions, tibial hinge component, resurfacing femur axial pin, poly spacer, tibial baseplate, tibial baseplate tapered screw, resurfacing femur, and resurfacing femur tapered screw.The surgeon placed an antibiotic spacer.No additional information regarding this adverse event has been reported.
 
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2023-00063, #3013450937-2023-00064, #3013450937-2023-00065, #3013450937-2023-00066, #3013450937-2023-00067, #3013450937-2023-00068, #3013450937-2023-00069, #3013450937-2023-00071.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The device history records and sterilization batch records were reviewed and there were no non-conformances identified that would have contributed to this adverse event.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4109: historical data analysis.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: investigation findings code updated to 3221: no findings available.H6: investigation conclusions code updated to 4315: cause not established.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2023 where the following eleos implants were explanted: two cemented stem extensions, tibial hinge component, resurfacing femur axial pin, poly spacer, tibial baseplate, tibial baseplate tapered screw, resurfacing femur, and resurfacing femur tapered screw.The surgeon placed an antibiotic spacer.No additional information regarding this adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
RESURFACING FEMUR, LEFT, SIZE 4
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
ORCHID OREGON
13963 fir st
oregon city OR 97045
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
9732645400
MDR Report Key16899529
Report Number3013450937-2023-00070
Device Sequence Number5372285
Product Code KRO
UDI-Device IdentifierB278RFL004E01M0
UDI-PublicB278RFL004E01M0
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberRF-L004E-01M
Device Catalogue NumberRF-L004E-01M
Device Lot NumberP220063-030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2023
Supplement Date Received by Manufacturer05/31/2023
Initial Report FDA Received Date05/09/2023
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25001208E, ELEOS TIBIAL POLY SPACER; P/N 25002114E, ELEOS RESURFACING FEMUR AXIAL PIN; P/N KSC01265E, ELEOS CEMENTED STEM EXTENSION; P/N KSC01530E, ELEOS STEM EXTENSION; P/N RF-TSCRW-01M, ELEOS RESURFACING FEMUR SCREW; P/N TB-2204E-01M, ELEOS TIBIAL BASEPLATE; P/N TB-TSCRW-01M, ELEOS TIBIAL BASEPLATE SCREW; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age65 YR
Patient SexMale
-
-