• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER Back to Search Results
Model Number 3300
Medical Device Problem Codes Defective Device (2588); Key or Button Unresponsive/not Working (4063)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this latitude programmer was returned to boston scientific without any reported allegations against its functionality or involvement with any adverse patient effects.Initial analysis identified a product performance issue and detailed testing was performed.
 
Additional Manufacturer Narrative
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; the programmer was able to interrogate a test pulse generator device several times, confirming successful communication between the programmer and device.The ecg and pacing system analyzer (psa) functions passed all testing.Additionally, the touchscreen was tested for functionality and calibration; no anomalies were observed.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.We have received reports of similar events where the latitude programmer screen became unresponsive to touch inputs during use.In most cases, the programmer required a system reboot to complete testing.Representatives from our post market quality assurance, manufacturing, and design assurance groups are actively investigating this unresponsive touchscreen behavior with the latitude programmer to determine the root cause.Based on our initial investigation, this behavior is most likely design-related.
 
Event or Problem Description
It was reported that this latitude programmer was returned to boston scientific without any reported allegations against its functionality or involvement with any adverse patient effects.Initial analysis identified a product performance issue and detailed testing was performed.
 
Additional Manufacturer Narrative
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; the programmer was able to interrogate a test pulse generator device several times, confirming successful communication between the programmer and device.The ecg and pacing system analyzer (psa) functions passed all testing.Additionally, the touchscreen was tested for functionality and calibration; no anomalies were observed.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.We have received reports of similar events where the latitude programmer screen became unresponsive to touch inputs during use.In most cases, the programmer required a system reboot to complete testing.Representatives from our post market quality assurance, manufacturing, and design assurance groups are actively investigating this unresponsive touchscreen behavior with the latitude programmer to determine the root cause.Based on our initial investigation, this behavior is most likely design-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Common Device Name
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS ELECTRONICA S DE RL DE CV
paseo del norte no
4640 guadalajara
zapopan, jal 45010
MX   45010
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16900178
Report Number2124215-2023-22802
Device Sequence Number6819350
Product Code OSR
UDI-Device Identifier00802526610110
UDI-Public00802526610110
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number3300
Device Catalogue Number3300
Device Lot Number010726
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2020
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/04/2023
Supplement Date Received by Manufacturer06/16/2023
Initial Report FDA Received Date05/09/2023
Supplement Report FDA Received Date06/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-