| Model Number |
3300 |
| Medical Device Problem Codes |
Defective Device (2588); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this latitude programmer was returned to boston scientific without any reported allegations against its functionality or involvement with any adverse patient effects.Initial analysis identified a product performance issue and detailed testing was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; the programmer was able to interrogate a test pulse generator device several times, confirming successful communication between the programmer and device.The ecg and pacing system analyzer (psa) functions passed all testing.Additionally, the touchscreen was tested for functionality and calibration; no anomalies were observed.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.We have received reports of similar events where the latitude programmer screen became unresponsive to touch inputs during use.In most cases, the programmer required a system reboot to complete testing.Representatives from our post market quality assurance, manufacturing, and design assurance groups are actively investigating this unresponsive touchscreen behavior with the latitude programmer to determine the root cause.Based on our initial investigation, this behavior is most likely design-related.
|
| |
|
Event or Problem Description
|
|
It was reported that this latitude programmer was returned to boston scientific without any reported allegations against its functionality or involvement with any adverse patient effects.Initial analysis identified a product performance issue and detailed testing was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.The programmer was powered on and the logfile was downloaded; the programmer was able to interrogate a test pulse generator device several times, confirming successful communication between the programmer and device.The ecg and pacing system analyzer (psa) functions passed all testing.Additionally, the touchscreen was tested for functionality and calibration; no anomalies were observed.Laboratory analysis was unable to determine how or why the touchscreen became unresponsive.No error or fault codes were recorded in the programmer's logfile and benchtop testing was unable to reproduce the behavior.We have received reports of similar events where the latitude programmer screen became unresponsive to touch inputs during use.In most cases, the programmer required a system reboot to complete testing.Representatives from our post market quality assurance, manufacturing, and design assurance groups are actively investigating this unresponsive touchscreen behavior with the latitude programmer to determine the root cause.Based on our initial investigation, this behavior is most likely design-related.
|
| |
|