| Model Number |
21625000 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4).10-may-2023: trended case device investigation concluded: the usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector.Investigation is still in progress; a final report will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
10-may-2023: on an unknown date ((b)(6) 2023 best estimate) back of the charger burned.Charger was pulled into the wall and the part where the cord plugs into the charger is where it was burned no patient harm or property damage reported.Unknown if original equipment was used.No further follow up information expected.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref# (b)(4) 10-may-2023: trended case device investigation concluded: the usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector investigation is still in progress; a final report will be submitted upon completion.07-jun-2023: technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: the chargers have been tested according to applicable safety standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.In case of moisture, sweat or dirt in the connector, there can be high temperatures inside the connector due to power dissipation when connected to the charger plug.There can be melting of the plastic around the connector and is visible to the naked eye.It is highly unlikely that the user would touch in case there appears to be melting or high temperatures.In the unlikely event of the user touching this overheated device, it can lead to superficial or minor injury that would in most cases require no medical intervention.There are no new clinical aspects (e.G., unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation 0378080 clin eval plan&rpt,wl acc and risk/benefit conclusion herein are sufficient.The risk investigation concluded: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register.Manufacturer's investigation is final.This is a final report.
|
| |
|
Event or Problem Description
|
|
10-may-2023: on an unknown date ((b)(6) 2023 best estimate) back of the charger burned.Charger was pulled into the wall and the part where the cord plugs into the charger is where it was burned no patient harm or property damage reported.Unknown if original equipment was used.No further follow up information expected.
|
| |
|