AVANOS MEDICAL INC. ON-Q CATHETERS-SILVERSOAKER Back to Search Results
Model Number PM010-A
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, a portion of a catheter used for an adductor canal placement (tka) was retained in a patient; the registered nurse (rn) attempted to remove the catheter by placing a warm compress over the catheter insertion site but was met with resistance.The patient had an appointment to follow up with the doctor on (b)(6) 2023.Additional information received 12apr2023 reported, the catheter was placed on (b)(6) 2023 and had been in place for 4 hours when the rn met with extreme resistance during removal.The physician was able to remove it in the office, the procedure required 4 stitches; the patient¿s current condition was not reported.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 08 may 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
All information reasonably known as of 05 jun 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The actual sample from the reported event was returned for evaluation, the silver soaker catheter was returned alone, (no pump or other components received).The returned portion of the catheter (infusion segment) measured approximately 6.5 inches.Visual examination of the device revealed multiple damages to the catheter including visible evidence of stretching and breaking in three specific areas.Area 1, noted as kinking, stretching was above the body segment (from the hub to the kink/stretch) measured approximately 9.5 inches.Area 2, noted as: stretching/ breaking ¿ occurred between the 2 and 1 marking on the body segment section.Area 3 noted as: kinking/ knot - was above the infusion segment.The reported event could/ be confirmed as reported; however, the root cause was undetermined.All information reasonably known as of 29 jun 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q CATHETERS-SILVERSOAKER
Common Device Name
CATHETER
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S. DE R.L. DE C.V.
ave noruega edificio d-1b
fraccionamiento rubio
tijuana b.c.
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16901951
Report Number2026095-2023-00058
Device Sequence Number17854365
Product Code BSO
UDI-Device Identifier00193494135331
UDI-Public00193494135331
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K051401
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPM010-A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2023
Supplement Date Received by Manufacturer05/10/2023
06/23/2023
Initial Report FDA Received Date05/10/2023
Supplement Report FDA Received Date06/05/2023
07/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROPIVACAINE .2%
Patient SexMale
-
-