| Model Number |
MFR100509 |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the field, prior to use, a micrusframe10 5mm x 9.7cm coil (mfr100509, 30537018) came out of the packaging frayed.There was no patient injury as the device was not clinically used.The procedure was successfully completed.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(6) updated sections on this medwatch: b4, g3, g6, h2, h6 and h10.Complaint conclusion: as reported by the field, prior to use, a micrusframe10 5mm x 9.7cm coil (mfr100509, 30537018) came out of the packaging frayed.There was no patient injury as the device was not clinically used.The procedure was successfully completed.No additional information is available.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30537018 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined.The instructions for use (ifu) indicate that upon receipt of the device, inspect the shipping carton and the unit for signs of damage.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Search Alerts/Recalls
|
|