| Model Number |
SUTURELASSO, 25 DEGREE TIGHT CURVE RIGHT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy the ring in front of the device broke off inside the patient.The broken piece was retrieved from patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Upon visual evaluation, it was noted that the distal tip tight curve right, suture lasso of the shaft was broken.No broken pieces were returned for investigation.The most likely cause of this condition is use error related to excessive force used to pry and leverage the device.Functional testing was not performed due to the damage to the device.
|
| |
|