• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.5/15; COVERED CORONARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PK PAPYRUS US 3.5/15; COVERED CORONARY STENT Back to Search Results
Model Number 434889
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
Three pk papyrus covered stents were used for treatment of a type iii perforation in the lcx.Stents #1 and #2 partially stripped off of balloon but were able to be removed.Stent #3 stripped off of the balloon in the coronary artery.It was ultimately retrieved via snare and removed, intact, from the body.The perforation sealed spontaneously with balloon tamponade alone.It is unsure which lot numbers relates to which stent.The other stents were reported separately.
 
Additional Manufacturer Narrative
Only the stent of the affected pk papyrus stent system was returned to biotronik and subjected to a detailed technical investigation.Images of the case such as angiographies could not be obtained.The provided clinical information (pk papyrus device registration form) and the product release documentation were reviewed to establish whether a device deficiency contributed to this event.The technical investigation revealed that the stent is severely deformed over its entire length, i.E.At one end several struts are bent and at the other end the stent is flat.The delivery system and the hemostatic valve used in the intervention were not provided for analysis.Review of the product release documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations no manufacturing related root cause could be identified.Considering the physicians description of the event, the root cause for the complaint event is most likely related to the handling during the procedure, i.E.The hemostatic valve was not fully opened which is warned against in the ifu.The treating physician rated the occurrence as both device- and procedure-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 3.5/15
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16906621
Report Number1028232-2023-02363
Device Sequence Number17507482
Product Code NIV
UDI-Device Identifier07640130447172
UDI-Public(01)07640130447172(17)2406
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2024
Device Model Number434889
Device Catalogue NumberSEE MODEL NO.
Device Lot Number05226644
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2023
Supplement Date Received by Manufacturer06/20/2023
Initial Report FDA Received Date05/10/2023
Supplement Report FDA Received Date06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age84 YR
Patient SexMale
-
-