| Model Number |
OPT942 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.Product background: the opt942 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices and is also compatible with f&p mr850 and 950 humidification system devices.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in taiwan reported via a fisher and paykel (f&p) field representative that the tubing of an opt942 optiflow + adult nasal cannula was found damaged before use.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Product background: the opt942 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices and is also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt942 optiflow + adult nasal cannula was not returned to f&p for evaluation.Our investigation is based on the information and photograph provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the tubing of the opt942 optiflow + adult nasal cannula was torn and stretched near the 3-way connector.Conclusion: our investigation was unable to determine the cause of the observed damage to the opt942 optiflow + adult nasal cannula.Based on our knowledge of the product the tubing was likely subjected to excessive force.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The user instructions which accompany the opt942 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state: "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in taiwan reported via a fisher and paykel (f&p) field representative that the tubing of an opt942 optiflow + adult nasal cannula was found damaged before use.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
|
| |
|