This device is classified as import for export, therefore 510k is not applicable.Model fb-8v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ocular fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ocular.In addition, our technician confirmed that the insertion flexible tube compressed (short in length), the insertion flexible tube buckled, the angle wire play, the operation channel (primary) leak, the ocular leak, the cfb (coherent fiber bundle) dirty, and the light guide cable dirty; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|