|
Investigation: visual inspection: the following observations were made during the visual inspection: piece of third-party catheter; deposits in the sprung reservoir.Permeability test: the test showed that the progav 2.0, the shuntassistant 2.0 and the sprung reservoir are permeable.Computer control test: to check the flow rate of the valves, the valves were tested on a miethke computer-controlled testing apparatus and passed the standard test.The flow of the test liquid was reduced step by step from 60 ml/h to 5 ml/h (in accordance with iso 7197).Distilled water was used as the test-liquid.The resulting pressure was measured.The computer-controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 2,48 cmh2o.This is within the specified tolerance of 3 cmh2o ± 3 cmh2o.Additionally, the shuntassistant 2.0 was tested according to standard procedure in the horizontal and vertical position of the valve.At a fixed opening pressure of 10 cmh2o in the vertical position, a pressure of 10 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant 2.0 had a pressure of 12,98 cmh2o.This is within the specified tolerance of 10 cmh2o +4/-2 cmh2o.The test, performed in the horizontal position of the valve at a reference flow of 20 ml/h showed that the shuntassistant 2.0 with a result of 4,56 cmh2o is operating not within the specified tolerance (0 to 4 cmh2o).According to our results, we can detect a presence of a slowed outflow in the horizontal position of the shuntassistant 2.0.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.The shuntassistant 2.0 is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: after dismantling of the valves, deposits were found in both valves.Results: for the sake of thoroughness and transparency, we would like to point out that an adjustment test and permeability test was already carried out after the revision at the hospital.Based on our investigation results, we have determined a slower outflow in the horizontal position of the shuntassistant 2.0 at the time of our investigation.Detected deposits can be named as the cause of the functional deviation.Existing deposits in the csf can temporarily impair the function of the valve and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further actions are required from our point of view.
|