MAUDE Adverse Event Report: COVIDIEN NANOCROSS 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN NANOCROSS 0.14 OTW PTA DILATATION CATHETER; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number AB14W030150150
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A physician was using a nanocross pta balloon for treatment of a calcified lesion with 90% stenosis in the proximal region of the su perficial femoral artery.There was mild tortuosity and severe calcification.A 6fr sheath and a non medtronic guidewire were used.Embolic protection was not used.A non-medtronic inflation device was used.The device was prepped as per the ifu with no issues identified.It was reported the patient had a cutdown with antegrade access, after inflation of balloon, balloon was deflated and began to be removed.Heard a loud crack and it was noticed the balloon had broken away from the catheter. patient had a cutdown already for an endart.Detachment occurred upon withdrawal into sheath, removed sheath and balloon catheter.Bits of the balloon remained on the wire.Used forceps to remove bits from wire.Exchanged wire and continued procedure.The device had passed through a previously deployed stent and resistance was encountered during advancement.No further patient injury reported for this event.
 
Additional Manufacturer Narrative
Product analysis the device was returned with the balloon detached at balloon bond and the distal end of the inner was detached approximately 32mm from the balloon bond.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOCROSS 0.14 OTW PTA DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16911243
Report Number2183870-2023-00155
Device Sequence Number6170425
Product Code LIT
UDI-Device Identifier00763000109868
UDI-Public00763000109868
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K132777
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/15/2025
Device Model NumberAB14W030150150
Device Catalogue NumberAB14W030150150
Device Lot NumberB478804
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2023
Supplement Date Received by Manufacturer06/21/2023
Initial Report FDA Received Date05/11/2023
Supplement Report FDA Received Date07/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-