|
The two devices subject of the events were returned to hufriedy for evaluation.One of the devices (y9) was manufactured in november 2009 and there was evidence of corrosion on the device and the handle was worn.The other device (120) was manufactured in january 2020 and the handle had visible abrasions.Each device was observed under magnification at the breakage point.The surfaces at the breakage point had indications of overload failure.The hufriedy instrument reprocessing guide includes the following language related to inspection of instruments when reprocessing (3.2), "inspect all instruments after the cleaning and rinsing step for corrosion, damaged surfaces, and impurities.Do not further use damaged instruments (for limitation of the numbers of re-use cycles, see 3.8 reusability section).If instruments are still visibly soiled, clean again.Re-sharpen instruments if necessary.Completely remove any residues from the sharpening process, such as metal residue or sharpening oil.In case sharpening is done, remember to repeat the cleaning and sterilization process." the instrument reprocessing guide further states (3.3), "light corrosion on the surface can be removed with hu-friedy penetrating oil (ips) or shine renew (ims-1455).If the corrosion cannot be completely eliminated, the instruments should be removed from use.Otherwise such corrosion could damage other instruments.After treating an instrument with shine renew or ips, the instrument must be cleaned and sterilized once more." hufriedy has contacted the customer for additional information including how the instruments were being used.To date, we have not received a response.A follow-up mdr will be submitted should additional information become available.
|