BERCHTOLD GMBH & CO. KG CHROMOPHARE OR LIGHT CEILING-MOUNTED; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number CH00000001
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by a nurse that a blue dome cover has fully separated.There were no reported injuries or adverse consequences.
 
Additional Manufacturer Narrative
It was reported that the blue dome cover separated from the f generation light.There was no patient involvement, no injuries, no adverse consequences or delay.Per rsk12227 revision v, the severity is s0.The exact root cause for the dome separation is unknown and return of the product to stryker is pending.Once further information is obtained, a supplemental will be filed.
 
Event or Problem Description
It was reported by a nurse that a blue dome cover has fully separated.There were no reported injuries or adverse consequences.
 
Additional Manufacturer Narrative
Nurse emailed in blue dome cover has fully separated salesforce case number (b)(4).As reported this issue was observed during morning cleaning setup.Picture provided confirms the issue of separation.Dhr of the affected product was reviewed and it evident that it passed inspection at the time of manufacturing.This is an old light head which was manufactured in year 2014.Manual 700000335 has proper instructions to replace the covers.Potential cause of this issue could be normal wear and tear due to age of the product.There was no patient involvement or adverse consequence reported.If further information is obtained a supplemental will be filed.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMOPHARE OR LIGHT CEILING-MOUNTED
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
GM  78532
Manufacturer (Section G)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
GM   78532
Manufacturer Contact
volker hornscheidt
ludwigstaler strasse 25
tuttlingen 78532
GM   78532
74611810
MDR Report Key16911687
Report Number0008010153-2023-00012
Device Sequence Number7205619
Product Code FSY
UDI-Device Identifier07613327296167
UDI-Public07613327296167
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCH00000001
Device Catalogue NumberCH00000001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2023
Supplement Date Received by Manufacturer04/18/2023
Initial Report FDA Received Date05/11/2023
Supplement Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-