| Model Number |
COMPR FT SCRW,3.5 MINI, 36MM |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that after an initial arthrodesis surgery for hallux rigidus on 22nd (b)(6)2022 the patient developed an granuloma formation.A surgical exploration was performed and a fragment of the screw was identified.An additional surgery was performed on (b)(6) 2023 to perform a fresection of the ganuloma.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint was confirmed.One unpackaged ar-8730-36hs batch 14695453 was received for evaluation.Visual inspection found that a small, tiny piece of the mini compression screw was received.Due to the device's size, it is impossible to perform measurements.However, the received piece color matches the finish: anodize teal color per drawing c16497 at revision 2.The most likely cause(s) for the reported failure can be a user error, improper bone preparation, or misaligned insertion.
|
| |
|