| Model Number |
SUTURE ANCHOR, BIO-CORKSCREW FT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder arthroscopy the anchor of the device broke off after the second rotation inside the patient.All broken parts were retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Visual evaluation, the device was returned without the implant.The investigation is based on the picture provided and noted that the corkscrew is damaged.The method used to prepare the bone and the bone quality encountered during the procedure was not provided.The most likely cause can be attributed to improper bone preparation or prying/leveraging the screw during insertion.
|
| |
|