| Model Number |
CR2 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer received a replacement device.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device does not power on when the lid is opened.As a result, defibrillation therapy may be delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that the cause of the reported issue was due to a faulty reed switch, designator sw5.The device was archived by stryker.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device does not power on when the lid is opened.As a result, defibrillation therapy may be delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Search Alerts/Recalls
|