• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD MEGA 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD MEGA 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0295-01
Medical Device Problem Codes Device Alarm System (1012); Leak/Splash (1354); Material Deformation (2976)
Health Effect - Clinical Codes Cardiac Arrest (1762); Thrombosis/Thrombus (4440); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2023
Type of Reportable Event Death
Additional Manufacturer Narrative
Initial reporter full name: (b)(6).Additional reporter information: (b)(6), director of risk management, patient safety, accreditation, regulation & licensing, and infection prevention / (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id #: (b)(4).
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, an iab rupture occurred with blood backing up into helium tubing.The customer was unable to provide further information regarding alarms or clinical situation.The critical patient was unable to be support with therapy and had to be resuscitated.Reported via medwatch report # mw5117071 received 03may2023: while patient was in intensive care unit blood was noted in the intra-aortic balloon pump helium line, which had been in place for 23 hours, requiring immediate shut off of device.Patient suffered cardiac arrest leading to thrombosis, inadequate perfusion, family decision to place on comfort care, and death.This report is for the iab.A separate report has been submitted for the cardiosave iabp under mfg report number 2249723-2023-02149.
 
Event or Problem Description
It was reported that during intra-aortic balloon (iab) therapy, an iab rupture occurred with blood backing up into helium tubing.The customer was unable to provide further information regarding alarms or clinical situation.The critical patient was unable to be support with therapy and had to be resuscitated.Reported via medwatch report #mw5117071 received 03may2023: while patient was in intensive care unit blood was noted in the intra-aortic balloon pump helium line, which had been in place for 23 hours, requiring immediate shut off of device.Patient suffered cardiac arrest leading to thrombosis, inadequate perfusion, family decision to place on comfort care, and death.Received on 10may2023: 82f presented to outside hospital, found to have nstemi, transported to (b)(6).Patient found to have left main disease and multivessel cad during (b)(6) 2023 cardiac cath; 40cc iabp placed for chest pain during pci.Patient transferred to icu and was in stable condition.No complications noted during placement or during pumping.On (b)(6) 2023, about 22 hours after insertion, the machine showed an ¿iab catheter restriction¿ alarm and when the rn assessed for a kink in the line, she visualized blood in the line, then immediately disabled iabp function.
 
Additional Manufacturer Narrative
Additional reporter: (b)(6).Reference complaint #(b)(4).
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 7.5FR. 40CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key16916007
Report Number2248146-2023-00326
Device Sequence Number20438782
Product Code DSP
UDI-Device Identifier10607567107974
UDI-Public10607567107974
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K120868
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2026
Device Model Number0684-00-0295-01
Device Catalogue Number0684-00-0293
Device Lot Number3000288436
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Supplement Date Received by Manufacturer05/10/2023
07/19/2023
Initial Report FDA Received Date05/11/2023
Supplement Report FDA Received Date05/30/2023
07/20/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARDIOSAVE / CH254265D7
Outcome Attributed to Adverse Event Death;
Patient Age82 YR
Patient SexFemale
-
-