• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.168.000S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E1: state & postal code (b)(6).E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed.No conclusion could be drawn, as no product was received.Based on the information available, it has been determined, that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from depuy synthes reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent the orif surgery using a clavicle plate l with the va locking screw for clavicle fracture.In the surgery, the driver shaft¿s tip did not match the locking screw head properly.The surgery was completed with a same size new screw.There was no surgical delay.It seems to be a defect in the recess of the locking screw head.No further information is available.This complaint involves one (1) device.This report is for one (1) va lockscr a, 2.7 he 2.4 self-tap l20 tan.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2b: additional product code: hrs.H3, h4, h6 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that va lockscr ã¸2.7 he 2.4 self-tap l20 tan was found the head bent.This can be related with the unable to assemble/disassemble allegation.A dimensional inspection for the va lockscr ã¸2.7 he 2.4 self-tap l20 tan was not performed as it is not applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the va lockscr ã¸2.7 he 2.4 self-tap l20 tan would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: rev.C (current and manufactured) dimensional inspection: n/a.Part # 04.211.020ts, lot # 8l91462, manufacturing site: werk selzach logistik , release to warehouse date: 22.Dec.2021, expiration date: 01.Dec.2026, supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Non-sterile part #04.211.020 non-sterile lot # 531p365 manufacturing site: werk selzach logistik release to warehouse date: 30.Nov.2021 supplier: (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16916233
Report Number8030965-2023-06046
Device Sequence Number2098860
Product Code HWC
UDI-Device Identifier10886982274557
UDI-Public(01)10886982274557
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103243
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/11/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.168.000S
Device Catalogue Number04.211.020TS
Device Lot Number8L91462
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/26/2023
Supplement Date Received by Manufacturer06/05/2023
07/13/2023
Initial Report FDA Received Date05/11/2023
Supplement Report FDA Received Date06/22/2023
07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - PLATES: CLAVICLE
Patient SexUnknown
-
-