ST PAUL DELTEC GRIPPER PLUS NEEDLES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-2961-24
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, that the customer infused medical fluid (physiological saline solution) into the patient through the product using a syringe.But backflow of medical fluid was not observed.So the customer removed the product from the patient, and passed medical fluid through the product with the syringe connected.However, at that time, medical fluid was unable to be drawn (only air bubble were drawn).No patient injury was reported.
 
Additional Manufacturer Narrative
This remediation mdr was generated under protocol b10010116, as a result of warning letter cms# 6(b)(6).No problems or issues were identified during this device history record review.One sample was received, decontaminated.The sample was received in used condition with its original package.During the functional testing, no occlusion was detected.The complaint was not confirmed.Root cause: based on the analysis conducted in the sample provided, the complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER PLUS NEEDLES
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16916265
Report Number3012307300-2023-05609
Device Sequence Number11303089
Product Code FPA
UDI-Device Identifier10610586027697
UDI-Public10610586027697
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2026
Device Model Number21-2961-24
Device Catalogue Number21-2961-24
Device Lot Number4142822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2023
Initial Report FDA Received Date05/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-