| Model Number |
21-2961-24 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, that the customer infused medical fluid (physiological saline solution) into the patient through the product using a syringe.But backflow of medical fluid was not observed.So the customer removed the product from the patient, and passed medical fluid through the product with the syringe connected.However, at that time, medical fluid was unable to be drawn (only air bubble were drawn).No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol b10010116, as a result of warning letter cms# 6(b)(6).No problems or issues were identified during this device history record review.One sample was received, decontaminated.The sample was received in used condition with its original package.During the functional testing, no occlusion was detected.The complaint was not confirmed.Root cause: based on the analysis conducted in the sample provided, the complaint was not confirmed.
|
| |
|