| Model Number |
4008-06 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Misfire (2532)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the anastoclip gc medium clips were misfiring and then coming out of the vessel hanging/loose after being deployed.There were no patient injury.Two devices were involved in the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Therefore, we could not conclusively determine the root cause of the reported incident.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.Due to similar issues reported we have previously initiated a capa to further investigate the reported issue.Please note that this is report 1 of 2.Report 1220948-2023-00082 was submitted for the remaining device related to this incident.
|
| |
|