| Model Number |
2025G-25 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
Numbness (2415)
|
| Date of Event |
04/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
"separated where the tubing meets the connector."the first occasion resulted in numbness in my extremities before i noticed the disconnection and could have been deadly had i gone to sleep without discovering the problem.".
|
| |
|
Additional Manufacturer Narrative
|
|
The tubing disconnect interrupted delivery supplemental oxygen to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The tubing disconnect interrupted delivery supplemental oxygen to patient.Complaint history reviewed.There has been 1 similar complaint in the previous 24 months for product group 12 - tubing/connector under defect code "bond joint separation/disconnection".Without rm, lot #, or pictures of complaint, root cause cannot be verified.Most likely root cause is little/no bonding agent present on tubing where separated from connector.Application of bonding agent is a manual process and variation of amount applied to tubing is inherit to the process.Defects are mitigated through inspections based on statistical techniques.Risk(rma-20017c): r36: oxygen delivery interrupted - tubing components not bonded together securely - s=6 o=1 rpn=6.Rpn < 25, therefore risk is acceptable.
|
| |
|
Event or Problem Description
|
|
"separated where the tubing meets the connector.""the first occasion resulted in numbness in my extremities before i noticed the disconnection and could
have been deadly had i gone to sleep without discovering the problem.".
|
| |
|