| Model Number |
16SOFT-HF-25 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
04/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
There was an interruption of therapy that could lead to hypoxia.
|
| |
|
Event or Problem Description
|
|
Kinking is preventing flow of oxygen. patient claims this is a life threatening issue as they are not able to get adequate oxygen.
|
| |
|
Event or Problem Description
|
|
Kinking is preventing flow of oxygen. patient claims this is a life threatening issue as they are not able to get adequate oxygen.
|
| |
|
Additional Manufacturer Narrative
|
|
There was an interruption of therapy that could lead to hypoxia.Complaint history reviewed.There have been no similar complaints for 1600hf-25 cannula in the previous 24 months.More information sent via text message and phone call from customer.Customer receives product from medical supplier as well as purchasing from amazon.Lot# 354817 was bent/kinked in package.Purchased from amazon.Sunmed does not use amazon as a distributor and cannot control the quality of product during transit.Most likely root cause of defects is product mishandling during transit.Risk(rma-20017a): r97: low oxygen delivery - cannula material deteriorates resulting in leaks prior to first use - s=6 o=1 rpn=6.Rpn < 25.R100: low oxygen delivery - cannula tubing deteriorates resulting in tubing collapse prior to first use - s=8 o=1 rpn=8.Rpn < 25, therefore risk is acceptable.Product was purchased through an unapproved distributor (amazon), issue will not be brought to carb.
|
| |
|