Model Number 436-28-008
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 04/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to dislocation.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
The reason for this revision surgery, it was reported as, revision surgery: "patient was dislocated due to the locking ring disengaging from the constrained liner.Head was removed and replaced" the actual length of in-vivo for the item(s) listed is unknown as the original surgery date was not provided or could be established.This investigation is limited in scope as only partial information was provided to djo surgical - austin for review.The revised item(s) was not returned for examination and the item and or lot number(s) was not provided.To adequately investigate this event, the part and lot number(s) are necessary.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported item(s) showed no present trends or on-going issues that are needing a review.There was no other information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER, ACETABULAR, FMP CONSTRAINED 28XMP8
Common Device Name
PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key16924890
Report Number1644408-2023-00572
Device Sequence Number17875146
Product Code KWZ
UDI-Device Identifier00888912085717
UDI-Public00888912085717
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K023794
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number436-28-008
Device Catalogue Number436-28-008
Device Lot NumberN/A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2023
Supplement Date Received by Manufacturer06/01/2023
Initial Report FDA Received Date05/12/2023
Supplement Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
436-28-000 LOT:.; 497-28-350 LOT:.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-