| Model Number |
48345T-01 |
| Medical Device Problem Codes |
Break (1069); Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the curved-tip stapler sureform reload 45, instrument involved with this complaint and completed the device evaluation.Failure analysis found the primary failure of exposed component to be related to the customer reported complaint.The reload was found to have the knife exposed within the knife track.The reload was partially fired based on in-house testing and was found to have a firing failure.A piece, measuring 0.281" x 0.044", was found missing at the distal part of the reload.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a blade was exposed on the stapler reload after firing.The procedure was completed with no reported injury.
|
| |
|