Model Number 48345T-01
Medical Device Problem Codes Break (1069); Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the curved-tip stapler sureform reload 45, instrument involved with this complaint and completed the device evaluation.Failure analysis found the primary failure of exposed component to be related to the customer reported complaint.The reload was found to have the knife exposed within the knife track.The reload was partially fired based on in-house testing and was found to have a firing failure.A piece, measuring 0.281" x 0.044", was found missing at the distal part of the reload.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, a blade was exposed on the stapler reload after firing.The procedure was completed with no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFORM
Common Device Name
SUREFORM 45 RELOAD BLACK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key16927421
Report Number2955842-2023-13390
Device Sequence Number17822651
Product Code GDW
UDI-Device Identifier10886874117115
UDI-Public(01)10886874117115(10)N/A
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K183224
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48345T-01
Device Catalogue Number48345T
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/19/2021
Initial Report FDA Received Date05/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient Age23 YR
Patient SexFemale
Patient Weight73 KG
-
-