Device evaluation summary: visual inspection of the opened device shows the hemostasis cap is missing and there is evidence of a crack in the connector.Reason for return was confirmed.Conclusion: complaint is confirmed for cracked cannula.Analysis found cracks on the cannula body; there is a potential this was caused by the supplier or during manufacturing.After analysis the cause of this complaint could not be determined.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received from august 2020 through august 2021 for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends.This regulatory report is being submitted as part of a retrospective review and remediation per (b)(4)as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received information that prior to use of a 96570 bio-medicus nextgen femoral venous cannula, the customer reported that there was a medium crack on the connector.Product was in the doctor's hand when seen, prior to use.It was confirmed that the device was not used on the patient.There was no patient involvement so no adverse effect occurred.
|