• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID KEELED M40; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID KEELED M40; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWE513
Medical Device Problem Codes Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 08/09/2017
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received a report from the national joint registry that contains unpublished collected data on the usage and the outcomes with the tornier shoulder system.The report details analysis provided for procedures performed between april 2012 ¿ may 2022.During the review of the report, it was identified that on (b)(6) 2017 a patient required revision surgery due " cuff insufficiency|aseptic loosening", which was not previously reported to the manufacturer.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history was not necessary because the implant affected by this complaint has been manufactured prior to 2018.As a result, this implant is expired and can no longer be put on the market or implanted.The complaint history review found no other complaint recorded for this lot number.Moreover, the post market surveillance reviews have not identified any signals for similar issues that would have challenged the efficacy and safety of this product.A review of the labeling did not indicate any abnormalities.The nature of the complaint doesn¿t necessitate a drawing review.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.H3 other text : device not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM ANATOMIC GLENOID KEELED M40
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16931454
Report Number3000931034-2023-00148
Device Sequence Number6925275
Product Code KWS
UDI-Device Identifier03700434022407
UDI-Public03700434022407
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K111902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberDWE513
Device Catalogue NumberDWE513
Device Lot NumberAA4909
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/17/2023
Initial Report FDA Received Date05/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-