| Catalog Number |
07027915190 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable testosterone g2 elecsys results from the cobas e 801 analytical unit serial number (b)(6).The initial result was 1029 ng/dl.The sample was repeated by beckman methodology and the result was 752.75 ng/dl.On (b)(6) 2023, the reagent cassette was exchanged and the assay recalibrated.The sample was repeated with a result of 1024 ng/dl.The questionable result was reported outside of the laboratory and the clinician questioned the results.It was not known which result was believed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
Sample from the patient was submitted for investigation.H3: na.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation testing of the patient sample confirmed the customer's results.No interference was found.The specific root cause could not be determined.
|
| |
|