• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS SMART PORT; PORT & CATHETER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS SMART PORT; PORT & CATHETER, Back to Search Results
Model Number CT75STSA
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Infiltration into Tissue (1931); Foreign Body In Patient (2687)
Date of Event 04/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
The patient had a smartport, from a tandem ato ct vtx sltit 7.5 att kit, placed for chemotherapy treatment due to pancreatic cancer.The port was placed left subclavian.4 months post placement, it was reported that the port "stopped working", so the patient went in for imaging.Under imaging, it could be seen that the catheter line had fractured and had migrated to the patient's pulmonary trunk into the pulmonary arteries.The patient then underwent explantation on (b)(6) 2023 and had a new port was inserted.The port body and attached portion of the catheter were removed at that time.This meant that there was a delay in chemotherapy access, potentially progressing the disease.A change in medication (anticoagulation) was also required, adding further risks to the patient.The patient still requires an additional procedure in a tertiary centre to remove the fractured line that has migrated.The replacement procedure had not been performed at the time of this report.
 
Additional Manufacturer Narrative
Returned for evaluation was one (1) smartport with catheter tubing attached.The fractured/detached end of the catheter tubing (17cm mark) is oval in appearance indicating that the fracture was due to flexural fatigue of the tubing.This is indicative of the flexing that occurs during pinch-off syndrome (between clavicle and first rib when catheter is place sub-clavian).The fracture location being ~14cm from the port body is a little longer distance than normal for pinch-off, however, this would make sense if place in a larger adult; patient is 47-year-old male, but weight was not provided.No manufacturing non-conformances were observed during complaint sample evaluation and many indicators point to pinch-off syndrome being likely root cause for this event.The customer's complaint description of catheter fracture is confirmed.The likely root cause for this event is pinch-off syndrome, which is cautioned in the device dfu as potential complication.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use}, which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT
Common Device Name
PORT & CATHETER,
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key16932237
Report Number1317056-2023-00061
Device Sequence Number17831734
Product Code LJT
UDI-Device IdentifierH787CT75STSA0
UDI-PublicH787CT75STSA0
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K101017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberCT75STSA
Device Catalogue NumberCT75STSA
Device Lot Number5732117
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2023
Supplement Date Received by Manufacturer07/29/2023
Initial Report FDA Received Date05/15/2023
Supplement Report FDA Received Date08/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
-
-