The healthcare professional reported that during a coil embolization procedure targeting a pulmonary arteriovenous malformation (avm), coil embolization was initiated from the drainer side beyond the shunt from the feeder.A microcatheter (unspecified brand) approached and coil embolization was started.The complaint coil, a 7mm x 33cm deltafill 18 (dlf180733 / 30778017) was used as the fifth coil, but resheathing was ¿impossible.¿ the entire system was removed.The complaint coil was replaced with another smaller sized coil and the procedure was completed with the implantation of fifteen (15) coils.A continuous flush was maintained through the microcatheter.There was no report of any negative patient impact.On 07-may-2023, additional information was received.The information indicated that there was no damage noted on the introducer sheath.There was no resistance reported at any time during the advancement of the complaint coil through the microcatheter.Neck modeling was not used.The complaint coil was not positioned at a relatively sharp angle to the microcatheter.The microcatheter was not removed from the target.On 14-may-2023, additional information was received.The information indicated that the reason the complaint coil needed to be resheathed was due to poor conformability; the coil diameter was not a match.Based on the additional information received on 14-may-2023, the event has been deemed usfda reportable under 21 cfr 803 with a classification of ¿malfunction.¿.
|
|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone: (b)(6).The initial reporter email address was not available / reported.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30778017) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the information available and without the complaint product available to be returned for analysis, the reported issue in the complaint cannot be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|