| Medical Device Problem Codes |
Overheating of Device (1437); Temperature Problem (3022); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has been requested for return.Further information will be provided when investigation is complete.
|
| |
|
Event or Problem Description
|
|
Patient claims the sensor burned through his shirt, no harm to the patient was reported.The sensor got heated and burned the patient's shirt.The patient removed the sensor and returned the kit early.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the sensor burnt through the patient's shirt.The device was returned for investigation and engineering evaluation was performed.Engineering evaluation was unable to replicate the reported complaint.The sensor passed visual inspection however failed functional inspection as the bluetooth test could not be performed.Corrosion identified on the analog printed circuit board likely prevented the device from completing function testing.The reported event is therefore inconclusive.No other problems were found.
|
| |
|