| Model Number |
650001DAR |
| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
04/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is in progress.The conclusions will be provided within the follow-up report once the investigation is completed.
|
| |
|
Event or Problem Description
|
|
The patient reported that the mattress overinflated and the patient fell from the mattress.There was no serious injury.The patient's sustained a laceration to the head which only required first aid measures.
|
| |
|
Additional Manufacturer Narrative
|
|
The automatt of the nimbus 4 mattress overinflated when a patient was laying on the mattress.As a result, the patient fell from the mattress.The patient sustained a cut on the head (not a serious injury).The involved mattress was returned to the manufacturer for evaluation.During the manufacturer's inspection, it was found that the membrane holder relocated blocking the airflow.The relocation of the holder caused the overinflation of the automatt and subsequently the patient's fall.The simulation confirmed the customer¿s allegation.In summary, the nimbus 4 mattress did not meet its specification since the automatt sensor pad overinflated with the load applied.The reported issue occurred when the device was used by the patient.The patient involved sustained a non-serious injury.The complaint was decided to be reportable due to the patient's fall which could have led to serious health consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the returned product was completed on (b)(6).Currently, all the gathered data are analysed.Once the final conclusions are available, the final report will be submitted.
|
| |
|