• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH LID-HANDPIECE/MODULAR FOR TRS; BATTERY, REPLACEMENT, RECHARGEABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH LID-HANDPIECE/MODULAR FOR TRS; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 05.001.231
Medical Device Problem Codes Mechanical Jam (2983); Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from canada that the lid device mode switch would not move to the settings on the handpiece device.During in-house engineering evaluation it was determined that the lid device would run in the locked position.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The lid device was evaluated and the reported condition that the mode switch would not move to the settings was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device was worn, the housing was bent/deformed, the leak tightness test failed, the device would run in the locked position, and there was component damage.It was further determined that the device failed pretest for general condition, leakage test, check for lid housing roundness, check in ¿lock¿ mode, check the function with power module and handpiece, and check incompatibility with the trs recon saw handpiece.The assignable root cause of these conditions was determined to be traced to the user, which is user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LID-HANDPIECE/MODULAR FOR TRS
Common Device Name
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16938024
Report Number8030965-2023-06173
Device Sequence Number17964662
Product Code MOQ
UDI-Device Identifier07611819474901
UDI-Public07611819474901
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2023
Supplement Date Received by Manufacturer05/16/2023
Initial Report FDA Received Date05/16/2023
Supplement Report FDA Received Date05/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-