| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Brain Injury (2219)
|
| Date of Event |
02/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the neurosurgical intervention of colloid cyst excision by ventricular endoscopy, the endoscope is held by a metal arm fixed to the intervention table in order to free the surgeon's hand.At the end of the operation, the clamping jaw used to hold the endoscope broke, causing movement of the endoscope which was in the patient's brain, causing bleeding from the ependymal veins and reducing visibility for the end.Intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
*this report was incorrectly filed under the importer mfr number - 2020550-2023-00079.The report was filed within the fda filing timeline.The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
During the investigation of the claimed article, a breakage of the holding arm could be detected, which is probably due to a reprocessing error or the resulting corrosion.Dark spots are visible from where the corrosion and the resulting weakening of the material could spread.In addition, impurities are visible, which probably caused micro cracks.According to the date of manufacture (02.2018), the article is 5 years old, during which time the cracks were able to spread and finally break.The ifu, therefore, points out that the articles should be checked for defects and damage before use.The event is filed under internal karl storz complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|