MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. G7 OSSEOTI MULTIHOLE 52MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. G7 OSSEOTI MULTIHOLE 52MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign: japan.Visual evaluation of the returned product found the sterile barrier (blister) exhibits damage visually consistent with thermal damage.Sterility has been breached.This complaint has been confirmed by review of the returned product.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the packaging operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the shape of the plastic inner tray was deformed.It looked as if it had been melted by heat.This implant was used however the inner tray can be returned.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: g3; h2; h3; h10.Visual evaluation of the returned product found the sterile barrier (blister) exhibits damage visually consistent with thermal damage.Review of the internal evaluation and tensile test results by the manufacturing site found the blister deformity does not affect the sterility.Review of complaint history found no additional related issues for the reported product.The condition of the device when it left zimmer biomet is considered non-conforming to specification.The root cause of the reported event is the packaging operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 52MM E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16942224
Report Number3007963827-2023-00130
Device Sequence Number20561824
Product Code LPH
UDI-Device Identifier00887868356025
UDI-Public(01)00887868356025(17)320713(10)65451219
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110010264
Device Lot Number65451219
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2023
Supplement Date Received by Manufacturer06/30/2023
Initial Report FDA Received Date05/16/2023
Supplement Report FDA Received Date07/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-