|
Manufacturer's investigation conclusion: the reported event of low voltage values while connected to the power module was confirmed via analysis of the submitted log file.The submitted system controller event log file contained approximately 47 days of data (on (b)(6) 2023 per the timestamp).Throughout the log file, the relative state of charge (rsoc) and bus voltage levels were observed to be lower than the expected voltage of approximately 14.4v while connected to the power module.The log file captured low voltage advisory alarms on (b)(6) 2023 from 21:28:36 to 21:29:06 due to the bus voltage dropping below the low voltage threshold while connected to the power module.In addition, the decreased voltage levels while connected to the power module resulted in backup battery faults, captured on (b)(6) 2023 at 21:29:10 and on (b)(6) 2023 at 10:30:38; however, the backup battery faults were not active long enough to activate an auditory/visual alarm.The alarms and decreased voltage levels did not affect the controller¿s ability to operate the pump at the set speed.The 14v patient cable was not returned for analysis.Additional information provided communicated that the issue with the system controller overheating resolved after exchanging the patient cable; however, this could not be confirmed as no products were returned for analysis.The submitted log file also did not contain any data that would indicate an overheating issue with the system controller.The root cause of the reported event could not be conclusively determined through this analysis.The lot number of the 14v patient cable was not reported with the event.Heartmate ii instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate ii patient handbook section 5 ¿ ¿alarms and troubleshooting¿ cover all alarms (auditory and visual), including low voltage alarms, and the actions to take if the alarms do not resolve.Heartmate ii instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate ii patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment, including their 14v patient cable.Additionally, heartmate ii patient handbook section 10 entitled ¿safety checklists¿, instructs users to regularly inspect their accessories ¿ including their patient cable ¿ for damage, and to replace any equipment that appears damaged or worn.The patient handbook and the instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|