Model Number FX414T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 04/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fx414t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt showed an overdrainage and adjustment difficulties the patient underwent a revision procedure.The complaint device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 8 years.Weight: 22 kilograms (kg).Height: 130 centimeters (cm).Gender: female.
 
Additional Manufacturer Narrative
Visual inspection: during the investigation, deposits on the connecting catheter was determined.Permeability test: a permeability test has shown that all components are permeable.During the permeability test were bloody liquid flushed out.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the both valves are not operating within the acceptable tolerance in either position.An accelerated outflow could be determined.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in both valves.To make the proteins / deposits in the shunt system more visible, they were colored using a staining solution.Results: in advance, we would like to point out that the product sent in was not submerged in liquid at the time of delivery.The testing of dry valves is only of limited value.The product properties can be influenced by dry residues of cerebrospinal fluid or blood.Despite this, we have examined the valve.Based on our examination results, we can determine, at the time of our examination, an accelerated discharge at both valves, in their respective relevant position, as well as a non-adjustability at the progav 2.0.The cause of the accelerated outflow and non-adjustability could be the deposits found.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 25
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key16943026
Report Number3004721439-2023-00126
Device Sequence Number6850835
Product Code JXG
UDI-Device Identifier04041906136567
UDI-Public4041906136567
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2022
Device Model NumberFX414T
Device Catalogue NumberFX414T
Device Lot Number20034508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2023
Initial Report FDA Received Date05/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 YR
Patient SexFemale
Patient Weight22 KG
-
-