• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (STAH); BIVENTRICULAR REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (STAH); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Catalog Number P/N 500101
Medical Device Problem Codes Break (1069); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The removed section of driveline will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event or Problem Description
The customer, a syncardia authorized distributor, reported that the patient observed air leaking from a cannula.
 
Additional Manufacturer Narrative
Device history record (dhr) review confirmed that the 70cc tah lot # 123283 was manufactured up to specifications and no anomalies were found.The provided photographic evidence showed a cannula tear near the distal end of the cpc connector barb.The tear was circumferential and spanned approximately 25% of the cannula.No functional testing is possible as the cannula continued to be in patient use after repair and no additional testing was needed as the tear was visible from the photographs.The reported leak was resolved after the tear was fixed with rescue tape.Failure investigation for this complaint confirmed the reported issue from examining the provided photographic evidence.The customer complaint was not replicated as the device was not returned; the root cause of the reported leak was determined to be a cannula tear.The cause of the cannula tear cannot be determined; however, historical investigations have identified the driveline/cannula junction as a point of increased stress.This is a known issue that is currently being addressed in capa-2022-0007.Failure investigation identified no other test failures or damage that could have contributed to the event.The cannula tear was repaired immediately per the provided instruction and no adverse impact to the patient was reported.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event or Problem Description
The customer, a syncardia authorized distributor, reported that the patient observed air leaking from a cannula.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (STAH)
Common Device Name
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
michael larsen
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key16943189
Report Number3003761017-2023-00072
Device Sequence Number15182475
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (Y/N)N
PMA/510(K) Number
P030011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2024
Device Catalogue NumberP/N 500101
Device Lot Number123283
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2023
Supplement Date Received by Manufacturer05/03/2023
Initial Report FDA Received Date05/16/2023
Supplement Report FDA Received Date04/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age31 YR
Patient SexMale
-
-