• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP, TSV, 4.1, 10, MTXF, MG,HA; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP, TSV, 4.1, 10, MTXF, MG,HA; DENTAL IMPLANT Back to Search Results
Model Number TSVT4H10
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.B3: date of event unknown / not provided.D6a: implant date unknown / not provided.D6b: explant date unknown / not provided.E1: email unknown / not provided.
 
Event or Problem Description
It was reported that the abutment portions was stuck to implant, unable to remove / detach.
 
Additional Manufacturer Narrative
Zimvie received one (1) tsvt4h10, (imp, tsv, 4.1, 10, mtxf, mg,ha) for evaluation (images are attached).Visual evaluation / functional test was performed, and mount was able to detach as intended.No malfunction.Dhr review was completed for the subject lot number 1256694.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1256694 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : does not disengage/release.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00541-haz, the most likely root cause determined from the investigation were missing or confusing instructions for use and excessive torque used to place restorative component or debris stuck in implant.Therefore, based on the available information, a device malfunction did not occur.The mount was able to detach as intended.No malfunction.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type h3: changed "yes" to "no" h6: entered evaluation codes h10: added manufacturer narrative h3 other text : device was not returned.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP, TSV, 4.1, 10, MTXF, MG,HA
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16944810
Report Number0002023141-2023-01368
Device Sequence Number12351454
Product Code DZE
UDI-Device Identifier00889024344198
UDI-Public(01)00889024344198(17)271202(10)1256694
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K133339
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberTSVT4H10
Device Catalogue NumberTSVT4H10
Device Lot Number1256694
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2023
Supplement Date Received by Manufacturer04/26/2023
Initial Report FDA Received Date05/17/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/03/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-