MEDOS INTERNATIONAL SÃ RL CH CORTICAL FIX X-TAB 7X50MM TI; ORTHOSIS, SPINAL PEDICLE FIXATION
|
Back to Search Results |
|
| Model Number |
186770150 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported that this was a spinal fusion for spinal stenosis on (b)(6) 2023.The tab of the screw broke before the torque wrench idled.The surgeon attempted final tightening of another screw with the torque wrench, but the torque wrench did not idle even though the surgeon tightened quite firmly, and it seemed like the torque wrench was about to break.There was no substitute for the torque wrench which was ready to use at the surgery.Therefore, the sales rep suggested that the surgeon stop the tightening of a set screw when the surgeon felt that the set screw was sufficiently tightened, because there was a risk of implant damage if the torque wrench was rotated until it idled.The sales rep could tell from the way the surgeon applied proper force and the sound emitted from the torque wrench that the surgeon had performed a sufficient tightening.The surgery was completed successfully within 30 minutes delay.The patient status/outcome is stable.No further information is available.This report is for one (1) (b)(4).This is report 2 of 2 for complaint viper2 final tightener handle.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: manufacture date added.H6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: a review of the receiving inspection (ri) for cortical fix x-tab 7x50mm ti was conducted identifying that lot number tbafck was released in one batch.Batch 1: lot qty of 264 units were released on 16 jul 2021 with no discrepancies supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.E1: initial reporter facility name and address added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|