• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION STERNALOCK BLU PLATE 4 HOLE L 100 DEG; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION STERNALOCK BLU PLATE 4 HOLE L 100 DEG; PLATE, FIXATION, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Fluid Discharge (2686)
Date of Event 03/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It is reported that the patient underwent a procedure and subsequently post operative drainage was observed from the drain attached to the patient.The following day, a computed tomography scan of the treatment site showed that the plate that had been attached to the sternum, had fallen off.Also, the screws had fallen into the chest cavity.In the sternal body and talar body, all screws on the one side had fallen out and were in the chest cavity.The patient remains under observation and no further intervention is currently planned.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).D10 ¿ medical products : iitem# 73-2623, lot# j7397832, sternalock blu plate 8 hole x; qty 2.Item# 73-2412, lot# unk, sternalock® blu system scrw 2.4x12mm cancelous lockng; qty 12.G2: foreign source: japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2023-00184; 0001032347-2023-00185; 0001032347-2023-00197; 0001032347-2023-00196; 0001032347-2023-00195; 0001032347-2023-00194; 0001032347-2023-00193; 0001032347-2023-00192; 0001032347-2023-00191; 0001032347-2023-00190; 001032347-2023-00189; 0001032347-2023-00188; 0001032347-2023-00187; 0001032347-2023-00186.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The reported event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERNALOCK BLU PLATE 4 HOLE L 100 DEG
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16945854
Report Number0001032347-2023-00183
Device Sequence Number2161855
Product Code HRS
UDI-Device Identifier00841036141042
UDI-Public(01)00841036141042(10)J7354048
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161896
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number73-2643
Device Lot NumberJ7354048
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Supplement Date Received by Manufacturer09/08/2023
Initial Report FDA Received Date05/17/2023
Supplement Report FDA Received Date09/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient SexUnknown
-
-