Model Number 0673 |
Device Problems
Difficult to Remove (1528); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/15/2023 |
Event Type
Injury
|
Event Description
|
It was reported that during routine follow-up of the device system, high right ventricular (rv) capture thresholds were observed.The rv lead was explanted and replaced.It was stated that there was some difficulty removing the lead during explant; however, the entire device was successfully removed.The lead is expected to be returned for analysis.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Visual inspection showed that the lead insulation was bunched, and the conductor coils were deformed approximately 320 mm from the terminal end.This is consistent with friction force when the lead was placed through a constricted area during removal.Testing was completed to assess lead electrical performance and inner insulation integrity.Measurements throughout these tests were within normal limits.A stylet insertion test was also performed and indicated no blockage in the lead lumen.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observation of high capture threshold; however, the observation of the bunched lead insulation and deformed conductor coils was consistent with the lead placed through a constricted area.
|
|
Event Description
|
It was reported that during routine follow-up of the device system, high right ventricular (rv) capture thresholds were observed.The rv lead was explanted and replaced.It was stated that there was some difficulty removing the lead during explant; however, the entire device was successfully removed.The lead was returned for analysis.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|