MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL DISTAL AUGMENT CEMENTED SIZE 7, 7+ 10 MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL DISTAL AUGMENT CEMENTED SIZE 7, 7+ 10 MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Radiolucent Lines (4848)
Date of Event 04/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the study patient developed a gradual onset of persistent pain and instability approximately one year post implantation.Radiographic images showed radiolucency surrounding the tibial implant and a triple phase bone scan displayed aseptic loosening.The patient underwent a revision where the femoral and tibial components were found to have minimal ingrowth.All components, except patella, were revised with a revision knee system.No intra-op complications reported.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not retrurned by the facility.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products & mdrs.42542007101, lot 64817993 , tibia fixed cemented left size e mdr: 0001822565-2023-01317.42555805110, lot 64847601, tibial augment cemented half block left lateral size ef 10mm mdr: 0001822565-2023-01318.42555805310, lot 64783331, tibial augment cemented half block left medial size ef 10mm mdr: 0001822565-2023-01319.42560007514, lot 64911499, stem extension tapered cemented 14 mm diameter +75mm mdr: 0001822565-2023-01320.42504606201, lot 64881021, femur cemented standard left size 7 mdr: 0001822565-2023-01321.42556606210, lot 64866990, femoral distal augment cemented size 7, 7+ 10mm mdr: 0001822565-2023-01323.42556806205, lot 64859470 , femoral posterior augment cemented size 7, 7+ 5mm mdr: 0001822565-2023-01324.42560007514, lot 64911499 , stem extension tapered cemented 14 mm diameter +75mm mdr: 0001822565-2023-01325.Additional associated products.42512600714, lot 64629122, articular surface fixed bearing (cps) left 14 mm unknown patella.110034355, lot k1905z22ba, refobacin bc r 1x40 us cement.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Medical records and radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the x-rays to be prior to event/symptoms, radiolucency, no infection, aseptic loosening around the tibial implant, and varus stress possibly causing pain.The complaint is confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL DISTAL AUGMENT CEMENTED SIZE 7, 7+ 10 MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16947988
Report Number0001822565-2023-01322
Device Sequence Number6356382
Product Code JWH
UDI-Device Identifier00889024556805
UDI-Public(01)00889024556805(17)301021(10)64882685
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K191625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42556606210
Device Lot Number64882685
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2023
Supplement Date Received by Manufacturer06/09/2023
Initial Report FDA Received Date05/17/2023
Supplement Report FDA Received Date06/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight109 KG
-
-