| Catalog Number |
305125 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: a device history record review was completed by our quality engineer team for provided material number 305125 and lot number 2244895.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3: see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd precisionglide¿ needle was blocked during the vaccine injection.This complaint was created to capture the 1st of 2 related incidents.The following information was provided by the initial reporter: "two different nurses used syringes from this box.Both reported that when they went to inject the vaccine into the patient, the vaccine would not go through the needle.".
|
| |
|
Search Alerts/Recalls
|