| Model Number |
H700488 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, communication issues with the generator resulted in a significant procedural delay.The generator was replaced and the procedure was completed with no adverse consequences to the patient.The ablation catheter and the mapping catheter were in place inside the heart and just before the rf delivery was to be performed, an error message of "pump not connected" suddenly displayed and the rf delivery could not be performed by the ablation catheter.There were no issues with the waveform or catheter display in ensite.The cool point pump and the irrigation pump communication cable were replaced, but with no resolution.The cool point pump was replaced, but with no resolution.The ampere generator was replaced with a device that was not on site resulting in a 1.5 hour delay while getting the replacement device.The issue was resolved and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One ampere¿ rf ablation generator mn h700488 and sn (b)(6) was received for evaluation.Based on the information provided to abbott and the investigation performed, root cause of the field reported event citing "pump not connected" was successfully isolated to a hardware failure on the controller board subassembly.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|
Search Alerts/Recalls
|