• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number H700488
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2023
Type of Reportable Event Malfunction
Event or Problem Description
During an atrial fibrillation procedure, communication issues with the generator resulted in a significant procedural delay.The generator was replaced and the procedure was completed with no adverse consequences to the patient.The ablation catheter and the mapping catheter were in place inside the heart and just before the rf delivery was to be performed, an error message of "pump not connected" suddenly displayed and the rf delivery could not be performed by the ablation catheter.There were no issues with the waveform or catheter display in ensite.The cool point pump and the irrigation pump communication cable were replaced, but with no resolution.The cool point pump was replaced, but with no resolution.The ampere generator was replaced with a device that was not on site resulting in a 1.5 hour delay while getting the replacement device.The issue was resolved and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
One ampere¿ rf ablation generator mn h700488 and sn (b)(6) was received for evaluation.Based on the information provided to abbott and the investigation performed, root cause of the field reported event citing "pump not connected" was successfully isolated to a hardware failure on the controller board subassembly.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16948269
Report Number2184149-2023-00110
Device Sequence Number6423240
Product Code LPB
UDI-Device Identifier05415067004888
UDI-Public05415067004888
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH700488
Device Catalogue NumberH700488
Device Lot Number8194863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2023
Initial Report FDA Received Date05/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-